Micke Method (2011)

Reference

Micke A, Lewis DF, Yu X. "Multichannel film dosimetry with nonuniformity correction." Med. Phys. 2011;38(5):2523-34.

Theory

The original multichannel method. For each pixel, find the dose \(D\) that minimizes the sum of squared residuals across all color channels:

\[D^* = \arg\min_D \sum_{c \in \{R,G,B\}} \left[ \text{pixel}_c - \frac{r_c + s_c D}{t_c + D} \right]^2\]

Implementation

chromadose uses vectorized Gauss-Newton refinement — not per-pixel scipy.optimize. This processes the entire image in ~10 Newton iterations, fully vectorized over all pixels simultaneously.

Usage

from chromadose.methods import MickeSolver

solver = MickeSolver()
dose_map = solver.solve(film, calibration)

Performance

  • 550x500 image: 0.14s
  • Uncertainty: from inter-channel disagreement (std of per-channel doses)